Auto Exhaust

PROTECTION FROM FUME BUILD-UP BY STORED CHEMICALS

After a day of work, when the fume hood is switched off for the night, chemicals stored in the Chemical Storage Cabinet* still emit fumes which can build up to significant levels overnight. The Auto Exhaust system enters your fume hood into a ‘sleep mode’. It turns on the fume hood periodically for short durations (duration set by us at time of installation) to vent any fume buildup overnight. Keeping you safe while still giving you great savings.

LabGuard’s Auto Exhaust allows you to not run blowers constantly. Depending on the kind of chemicals you are storing, Auto Exhaust let’s you run the blower for as less as 5 min/ hour, giving you complete protection and great savings.